A blanket layer of silicon can be added to a silicon substrate by a CVD process to achieve changes in the properties, like resistivity, type and defect density. This CVD process is called epi (or epitaxial) deposition. The thickness monitoring of the epi layer is part of the production process of making the epi wafers. This monitoring is executed on a sample basis, or only to confirm that the epi reactor is set up properly.
Semilab can offer non-contact optical solution for Epi thickness monitoring, depending on infrared reflection techniques.
Figure 1. Resistivity measurement flow
The Air-Gap CV (ACV) technology provides non-contact measurements for epi layers, or substrate resistivity that lowers cost by eliminating test wafer scrap associated with the traditional means of monitoring this specification. All types of the epi substrate can be monitored: p/p+, n/n+, p/p-, n/n+/p+, bare n or n surface.
ACV is conceptually similar to conventional Hg-probe CV or Schottky-CV techniques. The depletion width is determined by applying different voltages at the wafer surface and measuring the resulting capacitance. An air bearing is used to position the sensor very near to the wafer surface without touching. This non-destructive and non-contaminating nature of the method makes it possible to perform reliable measurements on fresh actual product wafers.
An integrated surface treatment chamber (PTC) is used to stabilize the wafer surface after the epi process by creating a uniform, electrically well-behaving native oxide layer.
Ozone (O3) is created by shining UV light into CDA
The wafer is exposed to O3 at elevated temperature (450-490 °C)
Result is a passivated, thus an electrically “well-behaved” native oxide:
PTC process does not affect the wafer characteristics in any way
For fresh wafers, the treatment stabilizes the surface
For old wafers, PTC improves the surface native oxide
Figure 3. Air-Gap CV
ACV delivers the industry’s first realtime, non-contact method to measure epi layer resistivity profiles in production. Fast feedback and the availability of the resistivity profile plots and wafer maps aid in troubleshooting for still higher control over epi processes.
100% non-contact, non-destructive measurements of epi layer resistivity by patented technology.
Applications:
The ACV systems provide 100% non-contact, non-destructive measurements for epi layer resistivity by patented technology. The physics behind the measurement is quite similar to Schottky-CV or Hg-probe, and thus the output is a traditional CV doping profile. The difference is that the electrode does not touch the wafer, resulting in substantial cost savings of monitor wafers.
The model includes complete automation, powerful software and excellent measurement repeatability.
Features and System specifications:
ACV-2200 | ACV-3000 | ACV-3100 | |
---|---|---|---|
Wafer Size | 100 to 200 mm | 300 mm | 200 and 300 mm |
End Effector | Paddle | Edge Grip | Paddle |
Loadport | Open Casette / SMIF | FOUP | FOUP / SMIF / Open casette |
Number of Loadports | 1 | 1 or 2 | 1 or 2 |
Factory Automation | SECS/GEM | Full | Full |
Request Info